PERFORMANCE |
All specifications are measured with a pulser having a pulse width of 2 ns FWHM, and a rise time of 150 ps. Where significant, the measurement is corrected for the rise times of the pulser, coaxial cable, and oscilloscope. The specifications are identical for OUTPUTS 1 and 2. |
GAIN |
Nominally 100 (50 to 150), non-inverting, into a 50-Ω output load. |
OUTPUT RISE TIME |
Typically 350 ps. |
BANDWIDTH |
(3 db) Typically 100 kHz to 1 GHz. |
NOISE |
<100 µV rms equivalent input noise over the full bandwidth. |
INPUTS AND OUTPUTS |
|
INPUT |
SMA input connector with 50-Ω input impedance (ac-coupled) and a 10-kΩ resistance to ground. Input protected to a maximum of ±1 V. |
OUTPUT 1 |
SMA output connector provides a linear output range from 0 to -1.75 V and a maximum output of -2 V into a 50-Ω load. Output impedance is 50-Ω, ac-coupled, and short-circuit protected. The unused output must be terminated with a 50-Ω load for proper operation of the other output. An optional SMA 50-Ω terminator is available for this purpose. |
OUTPUT 2 |
Indentical to OUTPUT 1. |
POWER |
Input power is supplied through a captive cable (lenght:3 m) with a standard preamplifier power connector (9-pin D type). Compatible with instruments that provide a preamplifier power connector. |
ELECTRICAL AND MECHANICAL |
POWER REQUIRED |
+24 V at 200 ma. Captive power cord with standard 9-pin D connector derives power from any instrument equipped with the standard preamplifier power plug (e.g., spectroscopy amplifiers, 4002P Portable Power Supply, 9307-P pico-TIMING Discriminator, etc.) |
WEIGHT |
Net 0.2 kg (0.4 lb)
Shipping 1.1 kg (2.4 lb) |
DIMENSIONS |
Alumimum housing9.5 X 6.4 x 2 cm (3.75 X 2.5 X 0.8 in.). |
OPTIONAL ACCESSORIES |
The Model 9306-P is designed for use with 50-Ω coaxial signal cables having SMA connectors. The desired optional cables and adapters can be selected from the ordering information. To avoid degradation of the 350-ps rise time through long signal cables, the total length of the input and output signal cables should be ≤1.7 m. If one of the outputs is not used, it must be loaded with an SMA50 terminator.
The Model 9306-P should be used with the Model 9307-P pico-TIMIMG™ Discriminator for optimum time resolution with ultra-fast detectors. |